Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis?
ABSTRACT: This study evaluated influence of addition of lead foil coupled to different types of digital receptors on image noise and on diagnosis of fractured instruments in teeth without and with root filling. Twenty-six molars had their root canals (n=67) cleaned and shaped and were divided into g...
Guardado en:
Autores principales: | , , , , , |
---|---|
Lenguaje: | English |
Publicado: |
Universidad de La Frontera. Facultad de Medicina
2021
|
Materias: | |
Acceso en línea: | http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2021000300734 |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
id |
oai:scielo:S0718-381X2021000300734 |
---|---|
record_format |
dspace |
spelling |
oai:scielo:S0718-381X20210003007342021-09-21Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis?Costa,Eliana DantasBrasil,Danieli MouraGaêta-Araujo,HugoBrito,Ana Caroline RamosOliveira-Santos,ChristianoFreitas,Deborah Queiroz endodontics instrumentation dentistry radiography ABSTRACT: This study evaluated influence of addition of lead foil coupled to different types of digital receptors on image noise and on diagnosis of fractured instruments in teeth without and with root filling. Twenty-six molars had their root canals (n=67) cleaned and shaped and were divided into groups with and without root filling and with and without fractured instruments. Images were acquired in PSP and CMOS sensor, with and without addition of a lead foil coupled to the digital receptors. Diagnostic values were obtained by the area under the ROC curve, and one-way ANOVA compared the groups. Noise in images was measured by the standard deviation of mean gray values in images with and without the lead foil, and compared by two-way ANOVA (a=0.05). Addition of lead foil did not interfere in the diagnostic accuracy for both digital systems (p>0.05) and did not decrease noise in the images (p>0.05). However, independently of the addition of the lead foil, the phosphor-storage plate presented higher noise compared to the sensor (p ?0.05). Addition of lead foil to the digital receptor did not interfere in image noise and with the diagnostic accuracy for detecting fracured endodontic instrument in ex-vivo conditions.info:eu-repo/semantics/openAccessUniversidad de La Frontera. Facultad de MedicinaInternational journal of odontostomatology v.15 n.3 20212021-09-01text/htmlhttp://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2021000300734en10.4067/S0718-381X2021000300734 |
institution |
Scielo Chile |
collection |
Scielo Chile |
language |
English |
topic |
endodontics instrumentation dentistry radiography |
spellingShingle |
endodontics instrumentation dentistry radiography Costa,Eliana Dantas Brasil,Danieli Moura Gaêta-Araujo,Hugo Brito,Ana Caroline Ramos Oliveira-Santos,Christiano Freitas,Deborah Queiroz Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
description |
ABSTRACT: This study evaluated influence of addition of lead foil coupled to different types of digital receptors on image noise and on diagnosis of fractured instruments in teeth without and with root filling. Twenty-six molars had their root canals (n=67) cleaned and shaped and were divided into groups with and without root filling and with and without fractured instruments. Images were acquired in PSP and CMOS sensor, with and without addition of a lead foil coupled to the digital receptors. Diagnostic values were obtained by the area under the ROC curve, and one-way ANOVA compared the groups. Noise in images was measured by the standard deviation of mean gray values in images with and without the lead foil, and compared by two-way ANOVA (a=0.05). Addition of lead foil did not interfere in the diagnostic accuracy for both digital systems (p>0.05) and did not decrease noise in the images (p>0.05). However, independently of the addition of the lead foil, the phosphor-storage plate presented higher noise compared to the sensor (p ?0.05). Addition of lead foil to the digital receptor did not interfere in image noise and with the diagnostic accuracy for detecting fracured endodontic instrument in ex-vivo conditions. |
author |
Costa,Eliana Dantas Brasil,Danieli Moura Gaêta-Araujo,Hugo Brito,Ana Caroline Ramos Oliveira-Santos,Christiano Freitas,Deborah Queiroz |
author_facet |
Costa,Eliana Dantas Brasil,Danieli Moura Gaêta-Araujo,Hugo Brito,Ana Caroline Ramos Oliveira-Santos,Christiano Freitas,Deborah Queiroz |
author_sort |
Costa,Eliana Dantas |
title |
Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
title_short |
Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
title_full |
Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
title_fullStr |
Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
title_full_unstemmed |
Can the Incorporation of a Lead Foil on the Backside of Digital Receptors Produce a Better Endodontic Diagnosis? |
title_sort |
can the incorporation of a lead foil on the backside of digital receptors produce a better endodontic diagnosis? |
publisher |
Universidad de La Frontera. Facultad de Medicina |
publishDate |
2021 |
url |
http://www.scielo.cl/scielo.php?script=sci_arttext&pid=S0718-381X2021000300734 |
work_keys_str_mv |
AT costaelianadantas cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis AT brasildanielimoura cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis AT gaetaaraujohugo cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis AT britoanacarolineramos cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis AT oliveirasantoschristiano cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis AT freitasdeborahqueiroz cantheincorporationofaleadfoilonthebacksideofdigitalreceptorsproduceabetterendodonticdiagnosis |
_version_ |
1718324146674335744 |