Bilateral comparison in Rockwell C hardness scale between INRiM and GUM
This bilateral comparison in HRC is conducted in order to confirm the accuracy claimed by National Institute of Metrological Research in Italy (INRiM) and Central Office of Measures in Poland (GUM). Also, this study compares the difference of measurement results between two modernized deadweight-typ...
Guardado en:
Autores principales: | , , |
---|---|
Formato: | article |
Lenguaje: | EN |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://doaj.org/article/2bb2799e35d74b38bc48a3b35cf31e0b |
Etiquetas: |
Agregar Etiqueta
Sin Etiquetas, Sea el primero en etiquetar este registro!
|
Sumario: | This bilateral comparison in HRC is conducted in order to confirm the accuracy claimed by National Institute of Metrological Research in Italy (INRiM) and Central Office of Measures in Poland (GUM). Also, this study compares the difference of measurement results between two modernized deadweight-type Rockwell's hardness standard machines (HSMs) from GUM and primary hardness standard machine (PHSM) from INRiM. The hardness blocks of about 20 HRC, 35 HRC, 45 HRC, 50 HRC, 60 HRC and 65 HRC, which all have uniformity less then ±0.4 HRC according to EN ISO 6508-3, were used in this comparison. |
---|