VALIDATION OF A HIGH PERFORMANCE LIQUID CHROMATOGRAPHY METHOD FOR CARBENDAZIM RESIDUES QUANTIFICATION IN TOMATOES

A high performance liquid chromatography method for carbendazim residues determination was adapted to the condition of our laboratory and validated for tomatoes samples. Carbendazim eluted at 3.13 minutes. The signal was linear over the concentration range 1 to 15 µg/ml with correlation coefficient...

Description complète

Enregistré dans:
Détails bibliographiques
Auteurs principaux: Veronica TANASA, Madalina DOLTU, Dorin SORA, Radu I. TANASA, Narcisa BABEANU
Format: article
Langue:EN
Publié: Stefan cel Mare University of Suceava 2018
Sujets:
Accès en ligne:https://doaj.org/article/6fc8695575d545b1826e98694542de05
Tags: Ajouter un tag
Pas de tags, Soyez le premier à ajouter un tag!