Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy

Objective: This study analyzed the correlation between the average segment width (ASW) and gamma passing rate according to the multi-leaf collimator (MLC) position error. Method: To evaluate the changes in the gamma passing rate according to the MLC position error, 21 volumetric modulated arc therap...