Axial length estimation error caused by hidden double-peak on partial coherence interferometry in an eye with epiretinal membrane: a case report
Yoshiyuki Kitaguchi, Shinsaku Yano, Fumi Gomi Department of Ophthalmology, Sumitomo Hospital, Osaka, Japan Abstract: Here we report a patient in whom there was a myopic shift after combined cataract surgery and pars plana vitrectomy against the epiretinal membrane, related to axial measurement est...