Triggered kV Imaging During Spine SBRT for Intrafraction Motion Management

Purpose: To monitor intrafraction motion during spine stereotactic body radiotherapy(SBRT) treatment delivery with readily available technology, we implemented triggered kV imaging using the on-board imager(OBI) of a modern medical linear accelerator with an advanced imaging package. Methods: Trigge...

Descripción completa

Guardado en:
Detalles Bibliográficos
Autores principales: Jihye Koo, Louis Nardella, Michael Degnan, Jacqueline Andreozzi, Hsiang-hsuan M. Yu, Jose Penagaricano, Peter A. S. Johnstone, Daniel Oliver, Kamran Ahmed, Stephen A. Rosenberg, Evan Wuthrick, Roberto Diaz, Vladimir Feygelman, Kujtim Latifi, Eduardo G. Moros, Gage Redler
Formato: article
Lenguaje:EN
Publicado: SAGE Publishing 2021